Under Pressure: The Reusability of the iCare Rebound Tonometer's Disposable Probes
Digital Document
Content type |
Content type
|
||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Collection(s) |
Collection(s)
|
||||||||||
Title |
Title
Title
Under Pressure: The Reusability of the iCare Rebound Tonometer's Disposable Probes
|
||||||||||
Resource Type |
Resource Type
|
||||||||||
Description |
Description
Background: This research study aims to evaluate the effect that disinfection of
tonometer probes has on intraocular pressure measurements using the ICare tonometer. Methods: Intraocular pressure was measured six times in forty eyes using the ICare tonometer and a new tonometer probe. The probe from each eye was then disinfected using Hydrogen Peroxide, then re-tested six times on the same eye that it had measured previously, and the results were compared to find any difference. Results: The average reading from the unused probes was 15.14 with a standard deviation of 3.84, whereas the disinfected probes had an average reading of 14.76 with a standard deviation of 3.76 Conclusion: There was no statistical or clinically significant difference found in the intraocular pressure measurements taken with the ICare Tonometer using new versus disinfected probes. |
||||||||||
Handle |
Handle
http://hdl.handle.net/2323/3912
|
||||||||||
Persons |
Persons
Author (aut): Matthews, Brian
Author (aut): Linsley, Chad
|
||||||||||
Genre |
Genre
|
||||||||||
Subject | |||||||||||
Origin Information |
Origin Information
|
||||||||||
Note |
Note
This paper is submitted in partial fulfillment of the requirements for the degree of Doctor of Optometry. 23 pages.
|
||||||||||
Related Item |
Related Item
|
||||||||||
Language |
Language
|
Language |
English
|
---|---|
Name |
bitstream_9833.pdf
|
MIME type |
application/pdf
|
File size |
5746628
|
Media Use | |
Authored on |
|