Measurement Of Off-Axis Refractive Error With The Grand Seiko WR-5100K Autorefractor.
Digital Document
Content type |
Content type
|
||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Collection(s) |
Collection(s)
|
||||||||||
Title |
Title
Title
Measurement Of Off-Axis Refractive Error With The Grand Seiko WR-5100K Autorefractor.
|
||||||||||
Resource Type |
Resource Type
|
||||||||||
Description |
Description
PURPOSE. To investigate ifthe Grand Seiko WR-5100K autorefractor will take measurements at positions of 10°, 20°, and 30° ofhorizontal eccentricity, and if it does, is
measurement error induced. METHODS. A series of distance measurements were taken at varying degrees of off-axis positions to the right and to the left of the line of sight at increments of 10°, 20°, and 30° and compared to the primary gaze baseline measurement. RESULTS. Readings were successfully taken at 10°, 20°, and 30° from primary position, with readings of increased eccentricity showing increased variability. CONCLUSION. Care needs to be taken to assure proper patient alignment so that measurements taken are accurate. |
||||||||||
Handle |
Handle
http://hdl.handle.net/2323/4052
|
||||||||||
Persons |
Persons
Author (aut): Dayringer, Matthew J.
|
||||||||||
Genre |
Genre
|
||||||||||
Subject |
Subject
|
||||||||||
Origin Information |
Origin Information
|
||||||||||
Note |
Note
This paper is submitted in partial fulfillment of the requirements for the degree of Doctor of Optometry. 14 pages.
|
||||||||||